Is there a simple way to perform label-free quantification of Host Cell Proteins?

1000 proteins quantified in e. coli mock lysate - HCP
Alphalyse
Jun 25. 2018

Question:

I find it difficult to get an accurate label-free quantification of Host Cell Proteins with LC-MS. It also seems that my data is difficult to reproduce. – Even if I perform multiple LC-MS runs for each sample.

Is there anything I can do to improve on the quantification and the reproducibility of the data?

 

Answer:

For quantitative proteomics experiments it has been normal procedure to include an isotope labeled peptide standard to obtain good HCP data. Another solution is the use of chemical labelling (like iTRAQ® or SILAC). Some people even use isotope labeled recombinant proteins. You do this to consider the digest efficiency on a full length protein.

All of the above methods are time-consuming – and they are not always possible to use.

So what can you do instead?

 

With the new SWATH® LC-MS technology, label-free quantification of HCPs is no longer a problem

The reasons for this are:

  • SWATH® analysis is a data-independent acquisition method. This results in very reproducible MS data – both for identification and quantification of HCPs.
  • Accurate absolute quantification of HCPs is now obtained by the addition of standard proteins. They are added in known amounts and are digested within the sample.
  • The SWATH® LC-MS is running at robust microflow rates. Therefore, there is no longer a problem with time-consuming nanoflow HPLC or carry-over.

 

Label-free quantification of thousands of proteins

In order to test the new LC-MS method, we applied SWATH® label free quantification to an E. coli cell lysate. With this method we could identify and quantify more than 1.000 proteins, up to a total sum of 969.158 ppm. The expected sum is 1.000.000 ppm – see graph below.

 

1000 proteins quantified in e. coli mock lysate - HCP

 

The analysis is so robust and reproducible that the standard deviation for individual proteins is below 20% in repeat experiments.

So to sum it up:
With the SWATH® technique it is now much easier to quantify proteins reproducibly. And in addition, it is done without isotope labeling.

 

 

More information on analysis of HCP:

The SWATH® technology is a form of data-independent analysis (DIA). You can learn more about the technology here: “What is SWATH”

Find out more on the identification and quantification of Host Cell Protein on the Alphalyse website: www.alphalyse.com/hcp

About Alphalyse

Alphalyse uses expertise in the field of protein chemistry and bioinformatics, combined with top-of-the-line mass spectrometry equipment, to provide protein analysis services globally.

Alphalyse aims to deliver high quality data and customer service to all of our clients in a fast and convenient way.

www.alphalyse.com

Newsletter Subscribe to The Alphalyse Newsletter and get more
information on protein analysis methods and applications
© 2019 Alphalyse, Inc.